EN554, EN285, EN15883, HTM2010, HTM2030, 21CFRpart11, AMS2750
Please visit Compatibility page for loggers & calibration hardware supported.
Security & 21CFR Part 11
- Login window with ID & password
- Password timeout & Renew
- Electronic Signatures on Reports
- Security Data is encrypted
- Complete Audit Trail detailing Change History
- Complete Audit Archive & Retrieve Changes Option
- Security Management Window
- Operator Profile Management
- Operator Timeout & Password length & timeout Controls
- Data Files Tamper Proofing
Calibration & Calibration Checking
- Select calibration file to store results
- Select channel groups for calibration
- Allow a test to use sensors calibrated in different groups
- User managed calibration stability and set point profile
- Automatic calibration option with automated stability analysis procedure
- Calibration progress monitor window with adjustable reference temperature for manual calibration
- Abort option
- Detailed comprehensive calibration reports stored & archived with management window
- Automatic management of calibration files. Test records are automatically assigned all relevant calibration reports for sensors used.
- Out of specification channels highlighted and when used in a test automatically set to defective
- Automatic management of calibration checking files. Test records are automatically assigned all relevant check reports for sensors used.
Test Specifications
- Allow user to create unlimited Test Specification files
- Allow new file to start with copy of existing one
- General Settings tab allows selection of:
- channel configuration file
- calibration file
- sterilization/disinfection temperature
- scan interval
- automatic stop options
- stage analysis (for HTM2010)
- continuous monitoring options
- selection of time format
- chart configuration
- data list configuration
- Calculations tab allows setup of calculations based on each row of data:
- Max per scan
- Min per scan
- Average per scan
- Max – Min per scan
- Max – Average per scan
- Average – Min per scan
- Min Lethality per scan
- Saturation temperature
- Saturation pressure
- Lethality tab allows configuration of lethality calculations:
- Base temperature & Z value
- Probe group
- Start lethality on temperature or stage
- End lethality on stage
- Temperature/Pressure correlation limit with reset option
- Option to use lethality config on time at temperature calculations
- Probe selection allowing location entry
- Setup of stages for manual entry during the test
- Setup of parameter specific to logger selected (eg TMI-ORION)
- Test Specification file is copied to test record for access historically as Test Detail
- Historic changes do not affect Test Specifications on Setup menu
- Test records have Test Setup report of Test Specification parameters
- Historic tests can be re-analysed after fully audited changes to historic test specification
- Channels can be set or unset to or from defective state
- Channels set as defective are not reported or included in any calculations
Test Specifications I-Calcs
- User configurable Interval Calculations
- User managed list window
- Allow I-Calc duplication
- Allow I-Calc copy to another Test Specification
- Select purpose of I-Calc:
- Chart line
- Stage or message insert
- Continuous limit monitoring
- Calculation on Interval
- Select group of probes
- Calculations available:
- Selectable groups
- Selectable Interval
- Selectable Limits
- Set IO on Limit
- Set Message on Limit
- Set Report colour on Limit
- Select where calculation appears
- Set insert stage event on Limit
- Interval Duration
- Time at Temperature
- Max probe value
- Min of Maxs.
- (Max – Min) Of Maxs.
- Min probe value
- Max of Mins.
- (Max – Min) Of Mins.
- Average probe value
- Max Average
- Min Average
- Fluctuation
- Max Fluctuation
- Max Deviation
- Max lethality
- Min lethality
- Lethality Deviation
- Allow entry of limit values:
- Target value
- Upper limit
- Lower limit
- Hysteresis
- Allow entry of actions:
- Message to show on chart and/or data list
- Insert message at offset from actual time I-Calc is triggered
- Highlight on failure
- Hide on chart or data list option
- Digital IO ON/OFF
- Display screen message
The Data List
- Highly detailed and configurable Data Listing option
- Displays stages and I-Calc actions & results
- Summary option allows data to be printed at intervals configured for each stage
- Data List can be dragged & resized
- Landscape or portrait option
- Prints data in configurable colors according to value
- If too many channels for one line, data is formatted to fit paper
- All calculations can be shown
- Electronic signature option
Calibration Files
- Allows multiple calibration files selected on user managed names
- Option to switch off calibration
- Select from names for calibration function
- Contains last calibrated date, time, pass/fail etc details for each channel
The Chart
- Chart is fully configurable
- Setup margins for printout
- Setup range for each unit
- Setup number of grid lines
- Setup trace thickness
- Real time chart all functions operational
- Displays stages and I-Calc actions
- Shows two vertical axis ranges, one temperature, the other non temperature (any units)
- Chart can be dragged & resized
- Chart displays trace ID at customizable interval
- Mouse pointer on chart shows values & time value at location
- Data line option shows values & time at sliding line
- Easy stage entry option at data line position
- Zoom into any area and the zoom in again
- Chart printout includes trace ID table
Company Name & System ID & Options
- User name heading all reports
- Configuration file ID tag for group users
- Selectable date format
Historical Test Storage & Retrieval
- All tests are stored in separate machine folders
- Each test has entry in Historic Test Index
- Test Index refresh option scans specified folder and all sub-folders for stored tests and creates new list
- Historic Test List can be configured to show any number of test data fields
- List can be filtered and sorted on date and time
- List can be configured for ‘tree view’ mode to group tests according to any field (eg hospital site)
- Next and previous buttons sequences through tests updating the chart window with next record
Data Management & Archiving
- Data Management window:
- Allows selection of multiple Tests, Calibration records, Calibration Check records, or user configuration files
- For user configuration files user can further select the type of files
- Report on space requirement
- Allows selection of drive or network location for copy target
- Stores an ID on the target
- Reports space available on target
- Allows selected files to be copied and or deleted
- Archive window allows:
- Target drive or network location selection
- Automatic archiving option at end of test or calibration
- Real time archiving as test is being logged
- Start button for archive process
- Shows progress of archive process
- When archiving only changed or new files are transferred
Machine Records Database
- User managed database of equipment or facilities tested with custom fields
- Select machine for each test
- Information automatically copied to test records
- Fields can be used in TQExpress Custom Forms
- Filter records
Calibration Files
- Allows multiple calibration files selected on user managed names
- Option to switch off calibration
- Select from names for calibration function
- Contains last calibrated date, time, pass/fail etc details for each channel
Test Equipment Database
Store all information on Test Equipment in database allowing selectable serial numbers.
- Test equipment information automatically copied to calibration reports
- Fields can be used in TQExpress Custom Forms
Channel Configurations
- Allows multiple channel configurations selected on user managed names
- Automatically configures most loggers
- Select Input Type
- Select Chart Colour
- Select Units, Decimal Places
- User managed channel names
Running a Test
- Selection of Test Specification
- Entry of Job reference
- Entry of cycle number
- Selection of machine or facility being tested
- Custom entries are available
- Most application functions available while test is logging
- Or multiple TQSoft™ programs can be run from same folder, one logging a test, the other being used to create reports
Real Time Display
- Real time display window shows all values as soon as received from logger
- Includes time above temperature, lethality, correlation etc.
- Shows time elapsed, current stage, elapsed stage time
The Test Merge Feature
- Data from different kinds of logger can be merged into a new test record
- Specify how time differences are handled
- Specify a Test Specification to use for the new merged test record
- Select data from any number of tests and select specific channels to merge
- Auto selection feature on matching sensor locations
The Stages Window
- Stages window allows editing of file containing all possible stages, test specification stage list, or test record stage list
- Allows alteration of ‘interval’ setting for each stage
- Allows stages to be added from list of all possible stages
- Allows stages to be removed
- Allows user to manually enter the times of stages
- In real time mode allows the next stage in the test specification list to be manually added
Security & 21CFR Part 11, Calibration & Calibration Checking, Test Specifications, Test Specifications I-Calcs and more with TQSoft™…