Version 8.0.2
IR2452 | TQAero I-Calcs modified by the graphic window now use hard coded ID rather than the user modifiable name. |
IR2453 | Fix for use of HPDAQ902A 16 channel card with KeySight 970A. |
Version 8.0.1
IR2449 | Correction to battery logger driver where non Madgetech loggers were being checked for claim status with meaningless messages. |
IR2450 | Correction to Test Spec editing where new test spec is duplicated. |
Version 8.0.0
IR2119 | TQSoft now allows for a set of calculations to be repeated on sections of a test. For example, min, max and average I-Calcs can be applied separately to heat up, sterilization and cooling stages, and three calculation sheets can display results for each stage.There is a sample test (Penetration) that illustrates this functionality. There is also a sample report Dist & Penetration that illustrates the reporting sheets. |
IR2120 | Groups can now be used to generate separate chart, data list and position diagram for each group in reports manager. |
IR2125 | The row calculations in the Data List are no longer restricted to temperature. If all channels are humidity, then calculations are for humidity. If a mix of channels, then first channel shown will determine the calculation type. |
IR2150 | Rate of change I-Calc can now insert stages precisely where a change of state occurs by back tracking to pinpoint a change of state. For example, when an increasing pressure is detected, the software can now backtrack to when pressure was either stable or stopped dropping, and insert a stage precisely. |
IR2149 | The audit trail now records use of logger and sensor calibration file at start of test or at start of CSV file import. |
IR2151 | During manual calibration the operator is alerted for bath setting change. |
IR2153 | Report Manager can now allow selection from calibration report or cal check reports lists, rather than just from cal reports linked to tests.Also included now is a generic calibration certificate template in Report Manager/Add Sections/TQSoft Reports. |
IR2170 | Modification for Graphtec driver for variable number of channels. |
IR2176 | Saturation pressure now uses same decimal place setting as a pressure channel (if one is setup) so that chart works correctly. |
IR2178 | Fix for situation where sterilization analysis hysteresis moves the sterilization start stage, nut code that deletes calculation stages does not activate. |
IR2180 | Historic tests selection window now includes a drop down for memory of user folder selections. |
IR2181 | SIKA baths now supported. |
IR2183 | Added support to read Grant Squirrel data files. |
IR2184 | When TQSoft analyzes a vacuum stage and the vacuum hold is perfectly held, the start of the hold is now correctly located. |
IR2185 | Added support for Gemini View2 and Plus2 data loggers. |
IR2186 | Correction to make sterilization/target/disinfection description correct in Test setup report. |
IR2187 | If users adds the folder ‘\TQSoft’ or ‘\TQAero’ to the archiving & synching target folder, software now strips that off since it will be duplicated. |
IR2192 | The test specification ‘test notes file’ field can now reference a USER MESSAGE that can be displayed to remind operator about validation tasks. |
IR2193 | Correction for reports where channel value at a particular stage is required. Non sequential channels were not handled correctly. |
IR2194 | Historical tests selection window enabling now no longer requires a software restart. |
IR2195 | The test specification ‘test notes file’ field can now reference a ‘reports sections’ (RLF) file so that reports sections will be auto added to a report when the test is added to a report. |
IR2196 | Correction for far east characters in the job reference field. |
IR2204 | Correction for Min Lethality I-Calc which did not check for full range channels returning zero. |
IR2205 | Patch for a program freeze caused by corrupt data in chart with xmax less than xmin. |
IR2207 | Correction to merge window where first test in the list would not expand channels. |
IR2211 | Max number of stages in a test increased to 2000 to allow for the extra requirements of multiple stage analysis. |
IR2216 | THP battery pressure logger is now supported. |
IR2217 | Multiple zoom views can now be setup from reports manager window. |
IR2226 | Added driver for thermal bath Ametek CTC155. |
IR2228 | Bath uncertainty no longer included in cal report uncertainty calculations, just logger and reference unit uncertainty. |
IR2238 | Correction for average I-Calc when a dependent stage is moved during analysis. |
IR2242 | When importing CSV file if units row is found the channel type is set to Type T regardless. Now it is set to type T only if the channel config channel type conflicts with units in CSV. |
IR2243 | In I-Calc spec tab its now possible to change a selected group of I-Calcs for settings stage start/end and all the hide settings. |
IR2249 | Report signatures can now be controlled with more options in setup options window. |
IR2251 | The cal check report will now display the date/time of the calibration that is being checked. (Will only work on cal checks performed with new version.) |
IR2252 | Now 60 characters are allowed for channel location field. But only 40 can be displayed in test setup report because of space. |
IR2253 | TQSoft now stores data in a new format TQD file which allows use of floating point numbers thus removing the resolution limitation of 32000 in version 7. |
IR2269 | Fix for advanced cal reference countdown timer which did not work in certain circumstance. |
IR2270 | The TeamViewer quick support link was updated. |
IR2271 | The advanced calibration report now does not repeat cal point operator information. |
IR2276 | Software no longer allows a corrupted baud rate to be used (without error message) but instead defaults back to 9600. |
IR2281 | In manual calibration adjust its now possible to edit the calibration time and date in usage section. Changes are audited. This is for transducers. |
IR2282 | The Job Reference now allows special character (for far east languages) but cycle number auto replaces special chars with underscore. |
IR2283 | Additional fix for IR2256 Anville logger scan times. |
IR2287 | Enhancements for legacy Kaye Digistrip unit including support for more channels and scan interval control for calibration. |
IR2289 | Correction for MODBUSTCP, Anville, Kaye loggers used in the ‘log in another TQSoft’. |
IR2290 | Test Equipment now has its own secured file that can be managed with data management options. |
IR2293 | Cal report out of range measurements now denoted by ‘—-’ along with dependent calculations.Rounding ‘away from zero’ now used in cal reports, and all calculations now use rounded values to avoid 0.01 discrepancies. |
IR2294 | Reference values now uses 3 decimal places. |
IR2302 | Corrections for text spacing for Spanish translation. |
IR2305 | For users with large number of I-Calcs that are related to the sterilizing/target/disinfection temperature, an adjustment to this temperature can now auto adjust I-Calc limits. |
IR2306 | Advanced cal reports now include a column shown maximum deviation of each channel during check phase. |
IR2307 | Correction for pressure calibration procedure where software asks user to adjust thermal bath! |
IR2308 | Correction for advanced pressure calibration where check points were using uncalibrated data. |
IR2309 | On start-up TQSoft 7 loads all the report references in Reports folder. With many reports this becomes very slow. Version 8 delays full loading until necessary. |
IR2310 | Support for T&D loggers TRZ file format and T&D imports now create audit trail entries. |
IR2313 | Export to CSV function is back which is a much faster export than the full export to Excel. |
IR2315 | A continuously monitored I-Calc can now have a hysteresis in time format (hh:mm:ss) which allows limit checking only after a delay from start of monitoring. |
IR2316 | Support for Graphtec firmware 1.28.Dead backup battery causing issues is addressed.Now allows 2 decimal places for thermocouples.More support for reading secure GDB data files from the unit. |
IR2317 | EA04/02 NIST amendment to Combined Uncertainty calculations. |
IR2318 | If Fluke loggers not connected it is now possible to nevertheless edit up to 60 channels. |
IR2319 | Modifications to Data List allow extra decimal place resolution on new file format to be shown on Data List up to 10 digits numbers. |
IR2320 | Scan intervals less than 1 second are not now auto adjusted. |
IR2322 | If a Test Spec notes file is setup to use an RLF (reports list) file as a pre-configured report, TQSoft will now auto import data at stop logging. |
IR2323 | Correction to eliminate risk of endless loop freeze when TQSoft searches for its data files. |
IR2324 | Support for humidity from dewpoint sensor. Its now possible to connect a dewpoint sensor to the logging unit and locate the sensor in the chamber. An I-Calc can be setup for each temperature and TQSoft calculates the humidity at that location without use of humidity sensors. |
IR2324 | This update involves the inclusion of ‘continuously monitored I-Calcs’ in the chart and data list. A feature many engineers will appreciate as it allows calculations like deviations and averages on groups to be put on data list (in place of the basic calculations tab that uses all channels.) |
IR2325 | Its now possible to mix THP and Gemini battery loggers with both time starting at zero and 24 hour format. |
IR2326 | Correction for battery logger interface when assigning a logger to a test spec position. |
IR2327 | Correction for THP loggers for estimated time set zero, where time was adjusted to 1 day limiting logging to 1 day. |
IR2331 | Correction for I-Calc limit signature where the signature window must appear on top of all other windows. |
IR2332 | Correction for advanced calibration root sum squared uncertainties. |
IR2334 | The Export to Excel/CSV functions no longer exports defective channels. |
IR2335 | If a Test Specification has no probe positions setup TQSoft will use the probe positions setup in the chamber database. |
IR2337 | Correction for data list showing analysis messages when showing continuous I-Calcs. |
IR2341 | The max deviation I-Calc will now show both the max and min measurements in the result stage. |
IR2343 | An I-Calc fail that uses the display message action will now make an audit trail entry of the message and test details. |
IR2344 | The start logging process now has a ‘More’ button on the final stage allowing the historic test spec to be edited, not affecting the original test spec. |
IR2347 | Correction for @TEXT and @DATA in TQExpress import data. |
IR2349 | Modification for use of T&D Graph application under Win 10 updates. Windows registry EnableLUA now needs to be set to zero. |
IR2350 | Merged tests can that contain mismatched scan times will now still calculate lethality’s correctly. |
IR2351 | Data List preview can now contain 250 pages. |
IR2355 | Using a comma in test equipment serial number field is now auto replaced by underscore because the comma is treated as a delimiter character in other parts of TQSoft. |
IR2356 | The delay in Start Logging window is extended to allow for days. |
IR2357 | Its now possible to enable a corrections window in a user profile so that fully audited corrections can be made to some fields prominent in the reports. All corrections are fully audited. |
IR2358 | Correction to the audit trail window in case its screen position is set outside the current screen. |
IR2359 | Removed specific reference to AMS2750 because TQAero enables compliance with a number of specs such as CQI-9, AFNOR etc. |
IR2360 | The INT:00:00 setting for a stage will no longer interrupt the interval of scans either side of it. |
IR2361 | In TQAero the parameter graphic window will now audit any parameter changes. |
IR2366 | Data management deleting files not allows a cancel if deletion fails for some reason. |
IR2369 | Correction for T&D data file import where locations were mistakenly assigned to labels. |
IR2372 | Driver for Isotech Millik reference unit. |
IR2375 | Export to Excel and CSV will no longer export defective probes. |
IR2380 | Sensor positions diagram now allows extra items to be added to the diagram, the name of each item can be specified. |
IR2382 | Correction for Fluke LAN option where a previously selected COM port will interfere with opening LAN. |
IR2383 | Chinese characters now allowed in the cycle number and job reference boxes for start test. |
IR2385 | KeySight 970A driver. |
IR2391 | Continuous I-Calcs included in Excel export with method to improve speed and summary option added to Excel export. |
IR2395 | Continuous monitoring is no longer interrupted by ‘no cal file’ alert. |
IR2398 | Correction for Calibration Files upload function if logger connection was not previously initiated. |
IR2399 | Correction for Chinese/Japanese units translation immediately after language switch. |
IR2402 | Correction for analysis of I-Calcs that reference a group containing no probes. The stage result is not created. |
IR2404 | TQSoft now supports Madgetech RFC1000 for distributed monitoring projects. TQSoft can now read battery status of MadgeTech loggers. |
IR2405 | Excel export is now able to use more than AZ column. |
IR2407 | Adjustment to data list landscape format so that the righthand vertical is always printed. |
IR2413 | Oversized diagnosis file AgilentError DLOG file is now automatically managed on startup. |
IR2414 | Fix for rare corruption of data that can cause chart freeze. |
IR2419 | Update for recent versions of Adobe DC Reader where it requires modified Registry references. |
IR2420 | Fix for blank page after sensor position diagram in PDF print function. |
IR2424 | Fix for lower limit checkbox being auto set after use of TQAero graphic. |
IR2425 | Added ability to remove marker for extra items in the sensor positions diagram. |