IR2531 | Fix for Inputs Display for measurements above 320.00. |
IR2532 | Fix attempt for rare but important problem found with data storage component. Its not clear that the fix will work but it will also make all data storage faster. |
IR2508 | For I-Calcs programmed with high and low limit, the stage message will now indicate fails on high or low limit. |
IR2535 | Fix attempt for USB communications drops on a customer’s system. The conclusion is that the software cannot recover communications in most instances. Various software components were tried to no avail. It seems that Fluke unit in certain environments will be affected by electrical noise with various effects ranging from temporary loss of comms to a complete freeze. LAN is being used instead. |
IR2537 | Test end data is now available for systems that log for many days. |
IR2539 | Fix for ISO15883 report template that was used on PC with regional settings decimal place comma and delimiter semi colon. |
IR2541 | Fix for 970A at start of test where TQSoft send a valid time setting to the unit, but the command causes an error. |